Factors affecting osteointegration and the use of early functional load to reduce the duration of treatment in dental implantation

Authors:

Kulakov A.A., Kasparov A.S., Porfenchuk D.A.

More about the authors

Journal: Stomatology. 2019;98(4): 107‑115

Read: 5230 times


To cite this article:

Kulakov AA, Kasparov AS, Porfenchuk DA. Factors affecting osteointegration and the use of early functional load to reduce the duration of treatment in dental implantation. Stomatology. 2019;98(4):107‑115. (In Russ.)
https://doi.org/10.17116/stomat201998041107

Recommended articles:
Treatment of infe­rtility in a patient with recu­rrent implantation failure. Russian Journal of Human Reproduction. 2026;(1):64-68

References:

  1. WHO. Informatsionnyi byulleten’ №318. Elektronnyi resurs. https:// www.who.int/mediacentre/factsheets/fs318/ru/.
  2. Avanesyan RA. Mediko-sotsiologicheskie, pedagogicheskie i metodicheskie aspekty povysheniya effektivnosti prepodavaniya stomatologii i dental’noi implantologii v vysshikh uchebnykh zavedeniyakh. Sovremennye problemy nauki i obrazovaniya. 2014;6:1216. (In Russ)
  3. Avanesyan RA, Perikova MG. Vliyanie bioaktivnogo pokrytiya vintovykh dental’nykh implantatov na sroki ikh osteointegratsii (eksperimental’noe Issledovanie). Mezhdunarodnyi studencheskii nauchnyi vestnik. 2017;2:44. (In Russ.)
  4. Parkhamovich SN. Dental’nye shtifty i shtiftovye protezy v stomatologii. Sovremennaya stomatologiya. 2012;1:22-23. (In Russ.)
  5. Poroshin AV, Shemonaev VI, Mikhal’chenko VF, Mikhal’chenko DV. Povyshenie effektivnosti osteointegratsii dental’nykh implantatov v eksperimente. Vestnik Volgogradskogo gosudarstvennogo meditsinskogo universiteta. 2015;3:55. (In Russ.)
  6. AAID (2011), Dental Implants Facts and Figures. https://www.aaidimplant.org/about/ Press_Room/Dental_ Implants_FAQ.html.
  7. Greenberg AM. Advanced dental implant placement techniques. J Istanb Univ Fac Dent. 2017;51(3, suppl 1):76-89. https://doi.org/10.17096/jiufd.17594
  8. Volkov AV. Sozdanie kostno-dental’nogo ekvivalenta — proryv v implantatsionnoi stomatologii? Kletochnaya transplantologiya i tkanevaya inzheneriya. 2006;1(1):28-29. (In Russ.)
  9. Tlupova AB. Sravnitel’naya kharakteristika dental’nogo i vnutrikostnogo implantata. Tsentral’nyi nauchnyi vestnik. 2017;2:(33):7-8. (In Russ.)
  10. Javed F, Almas K. Osseointegration of dental implants in patients undergoing bisphosphonate treatment: a literature review. J Periodontol. 2010;81(4): 479484. https://doi.org/10.1902/jop.2009.090587
  11. Khabiev KN. Vliyanie dizaina implantata na uvelichenie pervichnoi stabilizatsii. Dental Magazin. 2014;2(122):88-90. (In Russ.)
  12. Chong L, Khocht A, Suzuki JB, Gaughan J. Effect of implant design on initial stability of tapered implants. J Oral Implantol. 2009;35(3):130-135. https://doi.org/10.1563/1548-1336-35.3.130
  13. Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil. 2006;33(1):75-78. https://doi.org/10.1111/j.1365-2842.2006.01545.x
  14. Alghamdi HS. Methods to Improve Osseointegration of Dental Implants in Low Quality (Type-IV) Bone: An Overview. J Funct Biomater. 2018;9(1):1. https://doi.org/10.3390/jfb9010007
  15. Degidi M, Daprile G, Piattelli A. Determination of primary stability: a comparison of the surgeon’s perception and objective measurements. Int J Oral Maxillofac Implants. 2010;25(3):558-561. PMID: 20556255.
  16. Gottlow J, Barkarmo S, Sennerby L. An experimental comparison of two different clinically used implant designs and surfaces. Clin Implant Dent Relat Res. 2012;14(suppl 1):204-212. https://doi.org/10.1111/j.1708-8208.2012.00439.x
  17. Mullodzhanov GE, Ashurov GG. Analiz sovremennykh vzglyadov na protsessy integratsii dental’nykh implantatov v kostnuyu tkan’. Vestnik poslediplomnogo obrazovaniya v sfere zdravookhraneniya. 2016;1:71-85. (In Russ.)
  18. Elias CN, Oshida Y, Lima JH, Muller CA. Relationship between surface properties (roughness, wettability and morphology) of titanium and dental implant removal torque. J Mech Behav Biomed Mater. 2008;1(3):234-242. https://doi.org/10.1016/j.jmbbm.2007.12.002
  19. Palmquist A, Omar OM, Esposito M, Lausmaa J, Thomsen P. Titanium oral implants: surface characteristics, interface biology and clinical outcome. J R Soc Interface. 2010;7(suppl 5):515-527. https://doi.org/10.1098/rsif.2010.0118.focus
  20. Shemtov-Yona K, Rittel D. An Overview of the Mechanical Integrity of Dental Implants. BioMed Res Int. 2015. https://doi.org/10.1155/2015/547384
  21. Stadlinger B, Pilling E, Huhle M, Mai R, Bierbaum S, Scharnweber D, Kuhlisch E, Loukota R, Eckelt U. Evaluation of osseointegration of dental implants coated with collagen, chondroitin sulphate and BMP-4: an animal study. Int J Oral Maxillofac Surg. 2008;37(1):54-59. https://doi.org/10.1016/j.ijom.2007.05.024
  22. Suzuki M, Guimaraes MV, Marin C, Granato R, Gil JN, Coelho PG. Histomorphometric evaluation of alumina-blasted/acid-etched and thin ion beam-deposited bioceramic surfaces: an experimental study in dogs. J Oral Maxillofac. Surg J Am Assoc Oral Maxillofac Surg. 2009;67(3):602-607. https://doi.org/10.1016/j.joms.2008.08.021
  23. Tetè S, Mastrangelo F, Traini T, Vinci R, Sammartino G, Marenzi G, Gherlone E. A macro- and nanostructure evaluation of a novel dental implant. Implant Dent. 2008;17(3):309-320. https://doi.org/10.1111/prd.12211
  24. Coelho PG, Jimbo R, Tovar N, Bonfante EA. Osseointegration: hierarchical designing encompassing the macrometer, micrometer, and nanometer length scales. Dent Mater Publ Acad Dent Mater. 2015;31(1):37-52. https://doi.org/10.1016/j.dental.2014.10.007
  25. Le Guéhennec L, Soueidan A, Layrolle P, Amouriq Y. Surface treatments of titanium dental implants for rapid osseointegration. Dent Mater Publ Acad Dent Mater. 2007;23(7):844-854. https://doi.org/10.1016/j.dental.2006.06.025
  26. Esposito M, Ardebili Y, Worthington HV. Interventions for replacing missing teeth: different types of dental implants. Cochrane Database Syst Rev. 2014; 7:CD003815. https://doi.org/10.1002/14651858.CD003815.pub4
  27. Steigenga JT, al-Shammari KF, Nociti FH, Misch CE, Wang HL. Dental implant design and its relationship to long-term implant success. Implant Dent. 2003;12(4):306-317. PMID: 14752967.
  28. Pedreira De Oliveira D, Ottria L, Gargari M, Candotto V, Silvestre FJ, Lauritano D. Surface modification of titanium alloys for biomedical application: from macro to nano scale. J Biol Regul Homeost Agents. 2017;31(2, suppl 1): 221-232. PMID: 28691477.
  29. Starikova SL. Vliyanie anodnogo okisleniya na passivatsiyu implantata i proteza. Vestnik problem biologii i meditsiny. 2013;2(1):297-299. (In Russ.)
  30. Dos Santos MV, Elias CN, Cavalcanti Lima JH. The effects of superficial roughness and design on the primary stability of dental implants. Clin Implant Dent Relat Res. 2011;13(3):215-223. https://doi.org/10.1111/j.1708-8208.2009.00202.x
  31. Mishra SK, Kumar MA, Chowdhary R. Anodized dental implant surface. Indian J Dent Res Publ Indian Soc Dent Res. 2017;28(1):76-99. https://doi.org/10.4103/ijdr.IJDR_386_16
  32. Szmukler-Moncler S, Testori T, Bernard JP. Etched implants: a comparative surface analysis of four implant systems. J Biomed Mater Res B Appl Biomater. 2004;69(1):46-57. https://doi.org/10.1002/jbm.b.20021
  33. Jinno Y, Jimbo R, Tovar N, Teixeira HS, Witek L, Coelho PG. In Vivo Evaluation of Dual Acid-Etched and Grit-Blasted/Acid-Etched Implants With Identical Macrogeometry in High-Density Bone. Implant Dent. 2017;26(6):815-819. https://doi.org/10.1097/ID.0000000000000672
  34. Guida L, Oliva A, Basile MA, Giordano M, Nastri L, Annunziata M. Human gingival fibroblast functions are stimulated by oxidized nano-structured titanium surfaces. J Dent. 2013;41(10):900-907. https://doi.org/10.1016/j.jdent.2013.07.009
  35. Val JEMS, Gómez-Moreno G, Ruiz-Linares M, Prados-Frutos JC, Gehrke SA, Calvo-Guirado JL. Effects of surface treatment modification and implant design in implants placed crestal and subcrestally applying delayed loading protocol. J Craniofac Surg. 2017;28(2):552-558. https://doi.org/10.1097/SCS.0000000000003209
  36. Yao C, Slamovich EB, Webster TJ. Enhanced osteoblast functions on anodized titanium with nanotube-like structures. J Biomed Mater Res A. 2008;85(1):157-166. https://doi.org/10.1002/jbm.a.31551
  37. Kazieva IE, Sirak S, Perikova M. Vozmozhnosti atomno-silovoi mikroskopii pri otsenke poverkhnosti vintovykh dental’nykh implantatov. Sovremennye problemy nauki i obrazovaniya. 2013;2:36. (In Russ.)
  38. Kulakov AA, Grigor’yan AS, Arkhipov AV. Vliyanie razlichnykh sposobov modifikatsii poverkhnosti dental’nykh implantatov na ikh integratsionnyi potentsial. Stomatologiya. 2012;91(6):75-77. (In Russ.)
  39. Mendonça G, Mendonça DB, Aragão FJ, Cooper LF. Advancing dental implant surface technology — from micron- to nanotopography. Biomaterials. 2008;29(28):3822-3835. https://doi.org/10.1016/j.biomaterials.2008.05.012
  40. Jonge LT, Leeuwenburgh SC, Wolke JG, Jansen JA. Organic-inorganic surface modifications for titanium implant surfaces. Pharm Res. 2008;25(10):2357-2369. https://doi.org/10.1007/s11095-008-9617-092.
  41. Junker R, Dimakis A, Thoneick M, Jansen JA. Effects of implant surface coatings and composition on bone integration: a systematic review. Clin Oral Implants Res. 2009;20(suppl 4):185-206. https://doi.org/10.1111/j.1600-0501.2009.01777.x
  42. Prodanov L, Lamers E, Domanski M, Luttge R, Jansen JA, Walboomers XF. The effect of nanometric surface texture on bone contact to titanium implants in rabbit tibia. Biomaterials. 2013;34(12):2920-2927. https://doi.org/10.1016/j.biomaterials.2013.01.027
  43. Anitua E, Piñas L, Alkhraisat MH. Early marginal bone stability of dental implants placed in a transalveolarly augmented maxillary sinus: a controlled retrospective study of surface modification with calcium ions. Int J Implant Dent. 2017;3(1):49. https://doi.org/10.1186/s40729-017-0111-5
  44. Östman P-O, Wennerberg A, Ekestubbe A, Albrektsson T. Immediate occlusal loading of NanoTiteTM tapered implants: a prospective 1-year clinical and radiographic study. Clin Implant Dent Relat Res. 2013;15(6):809-818. https://doi.org/10.1111/j.1708-8208.2011.00437.x
  45. Farronato D, Mangano F, Briguglio F, Iorio-Siciliano V, Riccitiello F, Guarnieri R. Influence of Laser-Lok surface on immediate functional loading of implants in single-tooth replacement: a 2-year prospective clinical study. Int J Periodontics Restorative Dent. 2014;34(1):79-89. https://doi.org/10.11607/prd.1747
  46. Nevins M, Kim DM, Jun SH, Guze K, Schupbach P, Nevins ML. Histologic evidence of a connective tissue attachment to laser microgrooved abutments: a canine study. Int J Periodontics Restorative Dent. 2010;30(3):245-255. PMID: 20386781.
  47. Kashiwagi K, Tsuji T, Shiba K. Directional BMP-2 for functionalization of titanium surfaces. Biomaterials. 2009;30(6):1166-1175. https://doi.org/10.1016/j.biomaterials.2008.10.04095.
  48. Li Y, Lee IS, Cui FZ, Choi SH. The biocompatibility of nanostructured calcium phosphate coated on micro-arc oxidized titanium. Biomaterials. 2008;29(13):2025-2032. https://doi.org/10.1016/j.biomaterials.2008.01.009
  49. Krasnikov AV, Annikov VV. Eksperimental’naya aprobatsiya osteointegratsionnykh svoistv pokrytii implantatov s biodegradiruemoi plenkoi nanoagregatov flavonoidov. Agrarnyi nauchnyi zhurnal. 2016;12:8-13. (In Russ.)
  50. Bougas K Jimbo R, Jimbo R, Vandeweghe S, Hayashi M, Bryington M, Kozai Y, Schwartz-Filho HO, Tovar N, Adolfsson E, Ono D, Coelho PG, Wennerberg A. Bone apposition to laminin-1 coated implants: histologic and 3D evaluation. Int J Oral Maxillofac Surg. 2013;42(5):677-682. https://doi.org/10.1016/j.ijom.2012.11.008
  51. Hunziker EB, Enggist L, Küffer A, Buser D, Liu Y. Osseointegration: the slow delivery of BMP-2 enhances osteoinductivity. Bone. 2012;51(1):98-106. https://doi.org/10.1016/j.bone.2012.04.004
  52. Lyasnikova AV, Lepilin AV, Lyasnikov VN, Smirnov DA, Mostovaya OS. Kompleksnye issledovaniya fiziko-khimicheskikh i mediko-biologicheskikh svoistv antimikrobnykh biokompozitsionnykh pokrytii dental’nykh implantatov. Vestnik saratovskogo gosudarstvennogo tekhnicheskogo universiteta. 2010;1(44):83-91. (In Russ.)
  53. Sirak SV, Perikova MG, Kazieva IE. Opredelenie srokov osteointegratsii vintovykh dental’nykh implantatov s bioaktivnym bonitovym pokrytiem in vivo. Kubanskii nauchnyi meditsinskii vestnik. 2013;6(141):169-172. (In Russ.)
  54. Baier RE, Meyer DE. Implant surface preparation. Int J Maxillofac Implants. 1988;3:9-20. PMID: 3074954.
  55. Cochran DL, Schenk RK, Lussi A, Higginbottom FL, Buser D. Bone response to unloaded and loaded titanium implants with a sandblasted and acid-etched surface: a histometric study in the canine mandible. J Biomed Mater Res. 1998;40(1):1-11. PMID: 9511093.
  56. Wennerberg A, Hallgren C, Johansson C, Danelli S. A histomorphometric evaluation of screw-shaped implants each prepared with two surface roughnesses. Clin Oral Implants Res. 1998;9(1):11-19. PMID: 9590940.
  57. Coelho PG, Granato R, Marin C, Teixeira HS, Suzuki M, Valverde GB, Janal MN, Lilin T, Bonfante EA. The effect of different implant macrogeometries and surface treatment in early biomechanical fixation: an experimental study in dogs. J Mech Behav Biomed Mater. 2011;4(8):1974-1981. https://doi.org/10.1016/j.jmbbm.2011.06.016
  58. Shigehara S, Ohba S, Nakashima K, Takanashi Y, Asahina I. Immediate Loading of Dental Implants Inserted in Edentulous Maxillas and Mandibles: 5-Year Results of a Clinical Study. J Oral Implantol. 2015;41(6):701-705. https://doi.org/10.1563/aaid-joi-D-14-00018
  59. Donskikh DA. Narushenie osteointegratsii implantata. Byulleten’ Meditsinskikh Internet-Konferentsii. 2015;5(11):1443. (In Russ.)
  60. Sysa OA. Prichiny vozniknoveniya oslozhnenii posle implantatsii zubov. Byulleten’ meditsinskikh internet-konferentsii. 2015;5(11):1445. (In Russ.)
  61. Ashurov GG. Analiz sovremennykh vzglyadov na protsessy integratsii dental’nykh implantatov v kostnuyu tkan’. Vestnik poslediplomnogo obrazovaniya v sfere zdravookhraneniya. 2016;1:71-75. (In Russ.)
  62. Gusev OF, Shishkanov AV. Oshibki i oslozhneniya v dental’noi implantatsii. Elektronnyi sbornik nauchnykh trudov zdorov’e i obrazovanie v XXI veke. 2006;8(5):217-218. (In Russ.)
  63. Albrektsson T, Buser D, Sennerby L. Crestal bone loss and oral implants. Clin Implant Dent Relat Res. 2012;14(6):783-791. https://doi.org/10.1111/cid.12013
  64. Ata-Ali J, Ata-Ali F, Bagan L. A Classification Proposal for Peri-Implant Mucositis and Peri-Implantitis: A Critical Update. Open Dent J. 2015;9: 393395. https://doi.org/10.2174/1874210601509010393
  65. Padial-Molina M, Suarez F, Rios HF, Galindo-Moreno P, Wang HL. Guidelines for the diagnosis and treatment of peri-implant diseases. Int J Periodontics Restorative Dent. 2014;34(6):102-111. https://doi.org/10.11607/prd.1994
  66. Abolin AB, Zaitseva MYu, Afinogenova AG, Netyl’ko GI. Issledovanie osteokonduktivnykh svoistv granulirovannogo steklokeramicheskogo materiala «Biosit» v eksperimente. Travmatologiya i ortopediya Rossii. 2008;4(50): 6771. (In Russ)
  67. Utyuzh AS, Zagorskii VA, Zagorskii VV. Uprugo-napryazhennye sostoyaniya kostnykh struktur chelyustei i cherepa cheloveka. Simvol nauki. 2016;23(14):175-178. (In Russ.)
  68. Williams DF. On the mechanisms of biocompatibility. Biomaterials. 2008;29(20):2941-2953. https://doi.org/10.1016/j.biomaterials.2008.04.023
  69. Utyuzh AS, Zagorskii VV, Kristal’ EA, Mikhailova MV, Nefedova IV. Protezirovaniya patsientov s oporoi na implantaty pri polnoi vtorichnoi adentii i povyshennom rvotnom reflekse. Uspekhi sovremennoi nauki i obrazovaniya. 2016;4(8):72-76. (In Russ.)
  70. Ashuev ZhA. K voprosu obosnovaniya optimal’nykh srokov funktsional’nykh nagruzok pri zubnoi implantatsii. Meditsinskii Alfavit. 2012;2(7):30-32. (In Russ.)
  71. Borsari V, Fini M, Giavaresi G, Rimondini L, Consolo U, Chiusoli L, Salito A, Volpert A, Chiesa R, Giardino R. Osteointegration of titanium and hydroxyapatite rough surfaces in healthy and compromised cortical and trabecular bone: in vivo comparative study on young, aged, and estrogendeficient sheep. J Orthop Res Publ Orthop Res Soc. 2007;25(9):1250-1260. https://doi.org/10.1002/jor.20413
  72. Campos FE., Gomes JB, Marin C, Teixeira HS, Suzuki M, Witek L, ZanettaBarbosa D, Coelho PG. Effect of drilling dimension on implant placement torque and early osseointegration stages: an experimental study in dogs. J Oral Maxillofac Surg J. 2012;70(1):43-50. https://doi.org/10.1016/j.joms.2011.08.006
  73. Larsson C, Thomsen P, Lausmaa J, Rodahl M, Kasemo B, Ericson LE. Bone response to surface modified titanium implants: studies on electropolished implants with different oxide thicknesses and morphology. Biomaterials. 1994;15(13):1062-1074. PMID: 7888577.
  74. Larsson C, Thomsen P, Aronsson BO, Rodahl M, Lausmaa J, Kasemo B, Ericson LE. Bone response to surface-modified titanium implants: studies on the early tissue response to machined and electropolished implants with different oxide thicknesses. Biomaterials. 1996;17(6):605-616. PMID: 8652779.
  75. Saini M, Singh Y, Arora P, Arora V, Jain K. Implant biomaterials: A comprehensive review. World J Clin Cases. 2015;3(1):52-57. https://doi.org/10.12998/wjcc.v3.i1.52
  76. Gorbach EN, Emanov AA, Ovchinnikov EN, Kuznetsov VP, Fefelov AS, Gorgots VG, Borzunov DYu, Gubin AV. Osteointegratsiya innovatsionnykh kastomizirovannykh implantatov v trubchatoi kosti (eksperimental’noe issledovanie). Sovremennye tekhnologii v meditsine. 2017;9:1. (In Russ.) https://doi.org/10.17691/stm2017.9.1.09
  77. Zagorskii ZV. Dental’naya implantatsiya. Materialy i komponenty. Simvol Nauki. 2016;9:2. (In Russ.)
  78. Ivanov SYu, Solodkii VG, Muraev AA. Rossiiskaya sistema stomatologicheskikh implantatov LIKO-M: opyt pyatiletnego klinicheskogo ispol’zovaniya. Stomatologiya. 2013;92(6):53-55. PMID: 24429792. (In Russ.)
  79. Kirilova IA, Taranov OS, Podorozhnaya VT. Osteointegratsiya kompozitsionnykh kostno-keramicheskikh materialov m eksperimente. Khirurgiya pozvonochnika. 2014;4:80-87. (In Russ.)
  80. Simakhov RV, Grishin PO. Preimushchestva CLEAN POROUSTM novogo tekhnologicheskogo metoda obrabotki poverkhnosti dental’nykh implantatov. Sciencerise. 2015;2(7):61-68. (In Russ.)
  81. Ye L. Current dental implant design and its clinical importance. Hua Xi Kou Qiang Yi Xue Za Zhi Huaxi Kouqiang Yixue Zazhi. West China J Stomatol. 2017;35(1):18-28. https://doi.org/10.7518/hxkq.2017.01.003
  82. Valente ML da C, de Castro DT, Shimano AC, Lepri CP, dos Reis AC. Analysis of the influence of implant shape on primary stability using the correlation of multiple methods. Clin Oral Investig. 2015;19(8):1861-1866. https://doi.org/10.1007/s00784-015-1417-4
  83. Romanos GE, Ciornei G, Jucan A, Malmstrom H, Gupta B. In vitro assessment of primary stability of Straumann implant designs. Clin Implant Dent Relat Res. 2014;16(1):89-95. https://doi.org/10.1111/j.1708-8208.2012.00464.x
  84. Xu C, Wei Z, Liu N, Sun F, Chen H, Lin T, Zhang B, Tang T, Lu E. The effect of implant shape and screw pitch on microdamage in mandibular bone. Clin Implant Dent Relat Res. 2015;17(2):365-372. https://doi.org/10.1111/cid.12100
  85. Glauser R. Implants with an Oxidized Surface Placed Predominately in Soft Bone Quality and Subjected to Immediate Occlusal Loading: Results from an 11-Year Clinical Follow-Up. Clin Implant Dent Relat Res. 2016;18(3): 429438. https://doi.org/10.1111/cid.12327
  86. Schiegnitz E. Clinical and Radiological Long-Term Outcome of a Tapered Implant System with Special Emphasis on the Influence of Augmentation Procedures. Clin Implant Dent Relat Res. 2016;18(4):810-820. https://doi.org/10.1111/cid.12338
  87. Andrade CL, Carvalho MA, Cury AADB, Sotto-Maior BS. Biomechanical Effect of Prosthetic Connection and Implant Body Shape in Low-Quality Bone of Maxillary Posterior Single Implant-Supported Restorations. Int J Oral Maxillofac Implants. 2016;31(4):92-97. https://doi.org/10.11607/jomi.4133
  88. Da Silva EF, Pellizzer EP, Quinelli Mazaro JV, Garcia Júnior IR. Influence of the connector and implant design on the implant-tooth-connected prostheses. Clin Implant Dent Relat Res. 2010;12(3):254-262. https://doi.org/10.1111/j.1708-8208.2009.00161.x
  89. Rilo B, da Silva JL, Mora MJ, Santana U. Guidelines for occlusion strategy in implant-borne prostheses. A review. Int Dent J. 2008;58(3):139-145. PMID: 18630109.
  90. Nishioka RS, Vasconcellos LGO de, Melo Nishioka GN de. Comparative strain gauge analysis of external and internal hexagon, Morse taper, and influence of straight and offset implant configuration. Implant Dent. 2011;20(2):24-32. https://doi.org/10.1097/ID.0b013e318211fce8
  91. Tsuge T, Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009;28(4):373-381. PMID: 19721272.
  92. Khafizova FA, Khafizova RG, Khafizova IR. Ispol’zovanie ustroistva s «Pamyat’yu» formy dlya sozdaniya kostnogo lozha pod dental’nyi implantat. 2014. (In Russ.)
  93. Shaikhaliev AI, Kpacnov MC, Il’ina AP, Yamckova OV, Pybakova EYu, Cventckaya NV, Yamckova VP, Yamckov IA. Vliyanie khimicheskoi prirody implantatsionnykh materialov na protekanie regenerativnykh protsessov v kostnom lozhe. Biofizika. 2016;61(4):813-822. (In Russ.)
  94. Giro G, Tovar N, Marin C, Bonfante EA, Jimbo R, Suzuki M, Janal MN, Coelho PG. The effect of simplifying dental implant drilling sequence on osseointegration: an experimental study in dogs. Int J Biomater. 2013;2013: 230310. https://doi.org/10.1155/2013/230310
  95. Jimbo R, Andersson M, Vandeweghe S. Nano in implant dentistry. Int J Dent. 2014;314819. https://doi.org/10.1155/2014/314819
  96. Anitua E, Carda C, Andia I. A novel drilling procedure and subsequent bone autograft preparation: a technical note. Int J Oral Maxillofac Implants. 2007;22(1):138-145. PMID: 17340908.
  97. Sener BC, Al-Nawas B, Tegner A, Sagheb K, Berres M, Kämmerer PW, Wagner W. Effects of irrigation temperature on heat control in vitro at different drilling depths. Clin Oral Implants Res. 2009;20(3):294-298. https://doi.org/10.1111/j.1600-0501.2008.01643.x
  98. Yeniyol S, Jimbo R, Marin C, Tovar N, Janal Malvin N, Coelho Paulo G. The effect of drilling speed on early bone healing to oral implants. Oral Surg Oral Med Oral Pathol Oral Radiol. 2013;116(5):550-555. https://doi.org/10.1016/j.oooo.2013.07.001
  99. Stacchi C, Vercellotti T, Torelli L, Furlan F, Di Lenarda R. Changes in implant stability using different site preparation techniques: twist drills versus piezosurgery. A single-blinded, randomized, controlled clinical trial. Clin Implant Dent Relat Res. 2013;15(2):188-197. https://doi.org/10.1111/j.1708-8208.2011.00341.x
  100. Javed F, Romanos GE. The role of primary stability for successful immediate loading of dental implants. A literature review. J Dent. 2010;38(8):612-620. https://doi.org/10.1016/j.jdent.2010.05.013
  101. Khobkek DA, Uotson RM, Sizn LD. Rukovodstvo po dental’noi implantologii. M.: Medpress-inform; 2010. (In Russ.)
  102. Shokirov ShT, Ganiev AA, Zainutdinov MO. Osteoporoz V Dental’noi Implantologii. Printsipy Lecheniya I Profilaktiki (obzor). Zhurnal teoreticheskoi i klinicheskoi meditsiny. 2016;4:160-163. (In Russ.)
  103. Brignardello-Petersen R. Osteoporosis may result in a small increase in marginal bone loss around dental implants in osteoporotic women. J Am Dent Assoc. 2017;148(10):e143. https://doi.org/10.1016/j.adaj.2017.06.034
  104. Medeiros FCFL, Kudo GAH, Leme BG, Saraiva PP, Verri FR, Honório HM, Pellizzer EP, Santiago Junior JF. Dental implants in patients with osteoporosis: a systematic review with meta-analysis. Int J Oral Maxillofac Surg. 2018;47(4):480-491. https://doi.org/10.1016/j.ijom.2017.05.021
  105. Kulakov AA, Korolev VM, Karayan AS. Ispol’zovanie autokostnykh transplantatov s tsel’yu uvelicheniya al’veolyarnykh otrostkov i zameshcheniya kostnykh defektov chelyustei pri dental’noi implantatsii. Stomatologiya. 2007;86(2):30-35. PMID: 17828065. (In Russ.)
  106. Yanushevich OO, Kozlova MV, Mkrtumyan AM, Belyakova AS, Kozlova LS. Osobennosti dental’noi implantatsii u patsientov s osteoporozom. Lechenie i profilaktika. 2014;3(11):85-90. (In Russ.)
  107. Benevolenskaya LI, Lesnyak OM. Osteoporoz. Diagnostika, profilaktika i lechenie. Klinicheskie rekomendatsii. 2010. (In Russ.)
  108. Trezubov VN, Fadeev RA, Arutyunov DS. Sposob ob»ektivnoi otsenki sostoyaniya kostnoi tkani chelyustei po dannym ortopantomogramm. Institut stomatologii. 2006;1(30):38-41. (In Russ.)
  109. Mall N, Dhanasekar B, Aparna IN. Validation of implant stability: a measure of implant permanence. Indian J Dent Res Publ Indian Soc Dent Res. 2011;22(3):462-467. https://doi.org/10.4103/0970-9290.87071
  110. Luk’yanenko AA, Kazantseva IA. Opyt primeneniya rezonansno-chastotnogo metoda dlya otsenki stabil’nosti i osteointegratsii dental’nykh implantatov. Sovremennye problemy nauki i obrazovaniya. 2014;4:291. (In Russ.)
  111. Meredith N. Assessment of implant stability as a prognostic determinant. Int J Prosthodont. 1998;11(5):491-501. PMID: 9922740.
  112. Tatli U, Salimov F, Kürkcü M, Akoğlan M, Kurtoğlu C. Does cone beam computed tomography-derived bone density give predictable data about stability changes of immediately loaded implants?: A 1-year resonance frequency follow-up study. J Craniofac Surg. 2014;25(3):e293-299. https://doi.org/10.1097/SCS.0000000000000727
  113. Song Y-D, Jun S-H, Kwon J-J. Correlation between bone quality evaluated by cone-beam computerized tomography and implant primary stability. Int J Oral Maxillofac Implants. 2009;24(1):59-64. PMID: 19344026.
  114. Rodrigo D, Aracil L, Martin C, Sanz M. Diagnosis of implant stability and its impact on implant survival: a prospective case series study. Clin Oral Implants Res. 2010;21(3):255-261. https://doi.org/10.1111/j.1600-0501.2009.01820.x
  115. Winkler S, Morris HF, Spray JR. Stability of implants and natural teeth as determined by the Periotest over 60 months of function. J Oral Implantol. 2001;27(4):198-203. https://doi.org/10.1563/15481336(2001)027<0198:SOIANT>2.3.CO;2
  116. Kahraman S, Bal BT, Asar NV, Turkyilmaz I, Tözüm TF. Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self-tapping dental implants. J Oral Rehabil. 2009;36(10):755-761. https://doi.org/10.1111/j.1365-2842.2009.01990.x
  117. Snijders RS, Wijk AJ van, Lindeboom JA. A comparative study of the OsstellTM versus the Osstell MentorTM to evaluate implant stability in human cadaver mandibles. J Oral Rehabil. 2013;40(10):774-779. https://doi.org/10.1111/joor.12093
  118. Ushakov RV. Primenenie dental’nykh implantatov BoneTrust plus pri neposredstvennoi dental’noi implantatsii i nemedlennoi nagruzke. Stomatologiya dlya vsekh. 2013;2:18-21. (In Russ.)
  119. Kotenko KM. Effektivnost’ nemedlennoi implantatsii razlichnymi tipami implantatov. Sibirskii meditsinskii zhurnal (Irkutsk). 2011;103(4). (In Russ.)
  120. Nesterov AP, Litvinov SD, Nesterov AA. Neposredstvennaya dental’naya implantatsiya s ispol’zovaniem materiala LitAr. Institut stomatologii. 2006;3(32):120-121. (In Russ.)
  121. Polupan P. Implantatsiya: perezagruzka. Odnoetapnyi protokol i monolitnye implantaty. Problemy stomatologii. 2014;2. (In Russ.)
  122. Pommer B, Valkova V, Ubaidha Maheen C, Fürhauser L, Rausch-Fan X, Seeman R. Scientific Interests of 21st Century Clinical Oral Implant Research: Topical Trend Analysis. Clin Implant Dent Relat Res. 2016;18(4):850-856. https://doi.org/10.1111/cid.12371
  123. Testori T, Weinstein T, Scutellà F, Wang HL, Zucchelli G. Implant placement in the esthetic area: criteria for positioning single and multiple implants. Periodontol. 2000;77(1):176-196. https://doi.org/10.1111/prd.12211
  124. Cannizzaro G, Gastaldi G, Gherlone E, Vinci R, Loi I, Trullenque-Eriksson A, Esposito M. Two or three machined vs roughened surface dental implants loaded immediately supporting total fixed prostheses: 1-year results from a randomised controlled trial. Eur J Oral Implantol. 2017;10(3):279-291. PMID: 28944356.
  125. Crespi R, Capparé P, Crespi G, Lo Giudice G, Gastaldi G, Gherlone E. Dental Implants Placed in Periodontally Infected Sites in Humans. Clin Implant Dent Relat Res. 2017;19(1):131-139. https://doi.org/10.1111/cid.12425
  126. Merli M, Moscatelli M, Mariotti G, Piemontese M, Nieri M. Immediate versus early non-occlusal loading of dental implants placed flapless in partially edentulous patients: a 3-year randomized clinical trial. J Clin Periodontol. 2012;39(2):196-202. https://doi.org/10.1111/j.1600-051X.2011.01821.x
  127. He Y, Hasan I, Keilig L, Fischer D, Ziegler L, Abboud M, Wahl G, Bourauel C. Numerical investigation of bone remodelling around immediately loaded dental implants using sika deer (Cervus nippon) antlers as implant bed. Comput Methods Biomech Biomed Engin. 2018;21(4):359-369. https://doi.org/10.1080/10255842.2018
  128. Mundt T, Al Jaghsi A, Schwahn B, Hilgert J, Lucas C, Biffar R, Schwahn C, Heinemann F. Immediate versus delayed loading of strategic mini dental implants for the stabilization of partial removable dental prostheses: a patient cluster randomized, parallel-group 3-year trial. BMC Oral Health. 2016; 17(1):30. https://doi.org/10.1186/s12903-016-0259-z
  129. Pichotano EC, de Molon RS, Freitas de Paula LG, de Souza RV, Marcantonio EJr, Zandim-Barcelos DL. Early placement of dental implants in maxillary sinus grafted with leukocyte and platelet-rich fibrin (L-PRF) and deproteinized bovine bone mineral. J Oral Implantol. 2018. https://doi.org/10.1563/aaid-joi-D-17-00220
  130. Tallarico M, Meloni SM, Canullo L, Caneva M, Polizzi G. Five-Year Results of a Randomized Controlled Trial Comparing Patients Rehabilitated with Immediately Loaded Maxillary Cross-Arch Fixed Dental Prosthesis Supported by Four or Six Implants Placed Using Guided Surgery. Clin Implant Dent Relat Res. 2016;18(5):965-972. https://doi.org/10.1111/cid.12380
  131. Mura P. Immediate loading of tapered implants placed in postextraction sockets: retrospective analysis of the 5-year clinical outcome. Clin Implant Dent Relat Res. 2012;14(4):565-574. https://doi.org/10.1111/j.1708-8208.2010.00297.x
  132. Bogaerde LV, Pedretti G, Sennerby L, Meredith N. Immediate/Early function of Neoss implants placed in maxillas and posterior mandibles: an 18-month prospective case series study. Clin Implant Dent Relat Res. 2010; 12(suppl 1):83-94. https://doi.org/10.1111/j.1708-8208.2008.00135.x
  133. Pozzi A, Tallarico M, Moy PK. Immediate loading with a novel implant featured by variable-threaded geometry, internal conical connection and platform shifting: three-year results from a prospective cohort study. Eur J Oral Implantol. 2015;8(1):51-63. PMID: 25738179.
  134. Pozzi A, Mura P. Immediate Loading of Conical Connection Implants: Upto-2-Year Retrospective Clinical and Radiologic Study. Int J Oral Maxillofac Implants. 2016;31(1):142-152. https://doi.org/10.11607/jomi.4061
  135. Zhulev EN, Troshin VD, Aleksandrov AA. Otsenka kachestva zhizni bol’nykh s tserebrovaskulyarnoi patologiei i poterei zubov. Kubanskii nauchnyi meditsinskii vestnik. 2012;4:139-143. (In Russ.)
  136. Kan VV. Vliyanie protezirovaniya s oporoi na implantaty na kachestvo zhizni patsientov s polnoi i chastichnoi poterei zubov. Yakutskii meditsinskii zhurnal. 2015;1(49):64-66. (In Russ.)
  137. Suzuki M, Guimaraes MV, Marin C, Granato R, Gil JN, Coelho PG. Histomorphometric evaluation of alumina-blasted/acid-etched and thin ion beam-deposited bioceramic surfaces: an experimental study in dogs. J Oral Maxillofac. Surg J Am Assoc Oral Maxillofac Surg. 2009;67(3):602-607. https://doi.org/10.1016/j.joms.2008.08.021
  138. Kulakov AA, Ashuev ZhA. Neposredstvennaya implantatsiya i rol’ rannei funktsional’noi nagruzki na implantat (eksperimental’noe issledovanie). Stomatologiya. 2007;1:23-27. PMID: 17495806. (In Russ.)
  139. Romanos GE. Wound healing in immediately loaded implants. Periodontol. 2000;68(1):153-167. https://doi.org/10.1111/prd.12058
  140. Güncü GN, Tözüm TF, Güncü MB, Yamalik N, Tümer C, Karabulut E, Kilinç K. Myeloperoxidase as a measure of polymorphonuclear leukocyte response in inflammatory status around immediately and delayed loaded dental implants: a randomized controlled clinical trial. Clin Implant Dent Relat Res. 2008;10(1):30-39. PMID: 18335584.
  141. Hassan KS, Kassim A, Al Ogaly A-UR. A comparative evaluation of immediate dental implant with autogenous versus synthetic guided bone regeneration. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;106(5):8-15. https://doi.org/10.1016/j.tripleo.2008.07.015
  142. Huang H-L, Chang YY, Lin DJ, Li YF, Chen KT, Hsu JT. Initial stability and bone strain evaluation of the immediately loaded dental implant: an in vitro model study. Clin Oral Implants Res. 2011;22(7):691-698. https://doi.org/10.1111/j.1600-0501.2010.01983.x
  143. Cavallaro J, Greenstein B, Greenstein G. Clinical methodologies for achieving primary dental implant stability: the effects of alveolar bone density. J Am Dent Assoc. 2009;140(11):1366-1372. PMID: 19884393.
  144. Deli G, Petrone V, De Risi V, Tadic D, Zafiropoulos GG. Longitudinal implant stability measurements based on resonance frequency analysis after placement in healed or regenerated bone. J Oral Implantol. 2014;40(4)438447. https://doi.org/10.1563/AAID-JOI-D-12-00014
  145. Lachmann S, Laval JY, Axmann D, Weber H. Influence of implant geometry on primary insertion stability and simulated peri-implant bone loss: an in vitro study using resonance frequency analysis and damping capacity assessment. Int J Oral Maxillofac Implants. 2011;26(2):347-355. PMID: 21483888.
  146. Simunek A, Kopecka D, Brazda T, Strnad I, Capek L, Slezak R. Development of implant stability during early healing of immediately loaded implants. Int J Oral Maxillofac Implants. 2012;27(3):619-627. PMID: 22616056.
  147. Torsiglieri T, Raith S, Rau A, Deppe H, Hölzle F, Steiner T. Stability of edentulous, atrophic mandibles after insertion of different dental implants. A biomechanical study. J Cranio-Maxillofac Surg. 2015;43(5):616-623. https://doi.org/10.1016/j.jcms.2015.03.001